3 - 10
Section 3-3 - Receiving and Unpacking the Equipment
Section 3-4
Preparing for Installation
3-4-1
Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-4-2
Physical Inspection
3-4-2-1
System Voltage Settings
Verify that Docking Cart is set to the correct voltage.
The Voltage settings for the LOGIQ V5/LOGIQ V3 is found on a label to the right of the Power switch
and External I/O, on the rear of the system.
3-4-2-2
Video Formats
Check that the video format is set to the locally used video standard, NTSC or PAL.
3-4-3
EMI Protection
This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
WARNING
WARNING
Connecting a LOGIQ V5/LOGIQ V3 to the wrong voltage level will most likely destroy it.
Содержание LOGIQ V5
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks...
Страница 145: ......
Страница 169: ......
Страница 170: ...2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...