GE
LOGIQ V5/LOGIQ V3
D
IRECTION
5496012-100, R
EVISION
3
B
ASIC
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 31
4-3-13-4 Archiving Images
(cont’d)
10.)The status menu appears. When the backup/move has been complete, press
Next
.
NOTE:
If you need to insert the next media, a message appears providing you with the media label.
Label the media, then insert the next media and press OK.
11.)When the backup is complete, the completion wizard page appears. Press
Finish
.
All databases, presets and images should now be saved to removable media.
Figure 4-26 EZBackup Wizard 3