GE
LOGIQ V5/LOGIQ V3
D
IRECTION
5496012-100, R
EVISION
3
B
ASIC
S
ERVICE
M
ANUAL
4 - 20
Section 4-3 - General Procedure
4-3-9-2
Color Flow Mode Controls
Table 4-5
Color Flow Mode Controls
Control
Possible Bioeffect
Description/Benefit
Flow Selection
No
In the Lower Extremity Vein (LEV) and Abdominal
applications, you can quickly select the flow state via a
shortcut on the Color Flow Mode menu.
Gain
No
Gain amplifies the overall strength of echoes processed in
the Color Flow window or spectral Doppler timeline.
Scale (Velocity
Scale)
Yes
Increases/decreases the Scale on the color bar.
Wall Filter
No
Filters out low flow velocity signals. It helps get rid of
motion artifacts caused from breathing and other patient
motion.
Size/Position
Yes
Adjust size and position of the color window.
Invert (Color Invert) No
Lets you view blood flow from a different perspective, e.g.,
red away (negative velocities) and blue toward (positive
velocities). You can invert a real-time or frozen image.
NOTE: Invert reverses the color map, NOT the color PRF.
Baseline
No
Changes the Color Flow or Doppler spectrum baseline to
accommodate higher velocity blood flow. Minimizes
aliasing by displaying a greater range of forward flow with
respect to reverse flow, or vice versa.
Baseline adjusts the alias point. The default baseline is at
the midpoint of the color display and at the midpoint of the
color bar reference display.
Angle Steer
Yes
You can slant the ROI of the Color Flow linear image left
or right to get more information without moving the probe.
The Angle Steer function only applies to linear probes.
Accumulation
No
Accumulation enhances the flow in an image.
Available in Color Flow, PDI, and B Flow.
Color Flow Line
Density
Yes
Optimizes the Color Flow frame rate or spatial resolution
for the best possible color image.
Map
No
Allows you to select a specific color map. After you have
made your selection, the color bar displays the resultant
map.
Map Compress
No
Change the gradation of color map.
Threshold
No
Threshold assigns the gray scale level at which color
information stops.
Frame Average
No
Averages color frames.
Transparency Map
No
Brings out the tissue behind the color map.
Spatial Filter
No
Smooths out the color, makes it look less pixely.
Flash Suppression
No
Activates/deactivates Flash Suppression, a motion artifact
elimination process.
Packet Size
Yes
Controls the number of samples gathered for a single
color flow vector.
Sample Volume
Yes
Adjusts the size of the color flow doppler transmit wave (or
pulse) and size (or length). Lower setting gives better flow
resolution and a higher setting increases sensitivity.
Содержание LOGIQ V5
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks...
Страница 145: ......
Страница 169: ......
Страница 170: ...2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...