7 - 4
Section 7-2 - Gathering Trouble Data
Section 7-3
Screen Captures
There may be times when the operator or field engineer will want to capture a presentation on the
screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
7-3-1
Check and Record the Store Key Function
Check the function of the Store Key in the event that the customer may have made some custom
settings.
1.) Press
Utility
on the keyboard.
2.) Select
Connectivity
from the
Utilities
Menu.
3.) Select the
Button
tab on the
Connectivity
screen.
4.) In the
Physical Print Buttons
field, select Store.
The Connectivity/Buttons Screen will be displayed as shown in
Figure 7-2 on page 7-4
.
If Store key is not set to Whole Screen, as shown in
Figure 7-2 on page 7-4
, proceed to step 5 to record
the customer’s customized settings.
5.) In the Destinations section, record the service that is displayed.
6.) In the
Physical Print Buttons
section, record the parameters related to the service.
7-3-2
Setting the Store Key to Screen Capture
If the Store Key is not set to screen capture:
1.) While on the Connect screen, with the Buttons tab displayed, go to the
Destinations
list.
2.) From the list select
Copy To Dataflow
. Press [>>] to add the selection to the
Printflow View
section.
3.) Ensure that the
Physical Print Buttons
section for capture Area is set to Whole Screen,
secondary Capture and No Image Compression.
4.) The Store Key should now be set up for whole screen capture, sending the screens to the
image buffer (clipboard).
Figure 7-2 Define Store Key Operation
Содержание LOGIQ V5
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks...
Страница 145: ......
Страница 169: ......
Страница 170: ...2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...