GE
LOGIQ V5/LOGIQ V3
D
IRECTION
5496012-100, R
EVISION
3
B
ASIC
S
ERVICE
M
ANUAL
4 - 4
Section 4-3 - General Procedure
4-3-2-3
Check System Date and Time
A warning message “Please check the system date and time are correct” appears on the screen when
the system is powered on. This warning message appears for the possible reasons:
•
The system is not boot up for over 14 days.
•
The system time has been changed by 24 hours earlier than the current system time of last boot-up.
•
The BIOS time is changed by 24 hours earlier than the current system by resetting BIOS time,
replacing BIOS module or changing BIOS time.
This warning message is to remind the user to check the system date in case the system date and time
is incorrect.
Move the cursor to
OK
and press
Cursor
key on the control panel to select
OK
. The system enters
scanning mode.
Check the system date and time. If it is incorrect, follow below steps to reset the system date and time.
1.) Enter Utility -> System -> General -> Date/Time.
2.) Reset the system date and time.
3.) Select
Apply
and then select
OK
.
4.) Select
Save
.
Figure 4-3
Check system date and time message
Содержание LOGIQ V5
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks...
Страница 145: ......
Страница 169: ......
Страница 170: ...2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...