GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
6 - 8
Section 6-2 - Monitor Adjustments
Brightness
Contrast
Before making any adjustments, record the settings for
contrast and brightness. In a dimly lit room, adjust the
monitor to Brightness 50 and Contrast 40. Increase the
contrast until the left most block in the second row from
the bottom is just visible.All the remaining blocks in the
last 2 rows of the image should now be visible.Reset
the contrast and brightness to the recorded levels.
LN1 to LN18
Luminance for different grayscale levels. Usually, the
facility’s Biomed Engineer tests each gray level map on
the monitor from LN-1 to LN-18 with a light meter and
then graphs these on a scale to ensure that the curve
falls within the GSDF’s standard.
This can be performed with all three LOGIQ Totus
monitors with GSDF enabled. Generally, this is
performed yearly.
LN-1
LN-18
QC
Quality Control. Confirm that this image is clear and
crisp, without jagged edges or lines.
Table 6-3 Test Patterns (Continued)
Test
Pattern
Description
Corresponding Graphic
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......