LOGIQ Totus
Basic Service Manual
3 - 62
Section 3-7 - Setting Up Connectivity and Tips
4. Configure the fields on each the page.
Use the
Next
and
Previous
buttons to move through each page.
Next button
Previous button
5. Under
Verify Timeout (sec)
, select the amount of time (in seconds) after which the system will stop trying to establish a connection to the
service.
6. Select
Verify Connection
.
NOTE
: This action checks the connection status with the server and
the compatibility of the service configuration.
This is an example of a successful compatibility test. If the test is
unsuccessful, Connection and/or Compatibility will be red and text will
display indicating the issues.
7. Repeat these steps as needed to configure all of the services for the configured devices.
Table 3-26 To Configure Services for a Special Device (Continued)
Step
Corresponding Graphic
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......