LOGIQ T
OTUS
Basic Service Manual
8 - 168
Section 8-10 - Replacing Back End Parts
8-10-2-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
4.
Unscrew 1 screw to remove the SSD from the
UEPP-T
Steps
Corresponding Graphic