InSite
Use Insite to establish a connection between the Ultrasound system and the GE backoffice (InSite).
Insite includes Autoconnect which is enabled by default. When enabled and when the customer allows
outbound connections over port 443, the Ultrasound system will automatically connect to Insite so there
are no additional steps to perform. Manual configuration is needed in the following scenarios:
•
The customer does not allow outbound connections over port 443.
•
When password authentication is required.
•
When a proxy server is used.
•
When a VPN connection is used.
Perform this procedure to manually set up Insite (with Autoconnect disabled).
Table 3-30 To Manually Configure InSite (without Autoconnect)
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Special Devices
.
2. Select
InSite
.
3. Select
Autoconnect
and then select
Disable
.
4. Configure the fields on the page.
5. Select
Done
.
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......