GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting
7 - 59
7-7-4 Device Management (Device Mgmt) (continued)
Use Table 7-26 to troubleshoot Device Management (Device Mgmt) through the error codes. These
error codes will appear in the IoTAgentService.log log file. This log file can be found under D:\log.
NOTE:
Error codes may change over time.
Table 7-26 Errors for the Device Management Option
Error
Details
Solution
Activate
Agent service config file is corrupted.
Please contact your GE HealthCare
service personnel.
Agent side config
file(IoTAgentConfig.json) is
corrupted.
Contact service for enabling re-
registration.
Activation failed. Missing input Parameter.
(Error code: 000104)
Invalid Account Email has been
entered.
Enter valid Account Email.
Activation failed. Please contact your GE
HealthCare service personnel. (Error code:
401)
Invalid Registration Key has been
entered.
Enter valid Registration Key.
Activation failed.
Invalid Server URL has been entered.
Enter valid Server URL.
Activation failed.
Invalid Proxy server or Proxy port has
been entered.
Enter valid Proxy server and Proxy
port.
Activation failed. Please contact your GE
HealthCare service personnel. (Error code:
403)
After successful registration again
perform register.
Contact service for enabling re-
registration.
Invalid fleet account email format.
Activation failed.
Invalid or empty Account Email format
field.
Enter valid Account Email.
Invalid proxy port. Activation failed.
Enter string value in Proxy port field.
Enter valid Proxy port.
Agent service config file is corrupted.
Please contact your GE HealthCare
service personnel.
Empty Registration Key field.
Enter valid Registration Key.
Agent service config file is corrupted.
Please contact your GE HealthCare
service personnel.
Empty Server URL field.
Enter valid Server URL.
Invalid server certificate.
Please contact your GE HealthCare
service personnel.
Invalid server certificate
Install valid certificate using Device
Connect Service Certificate.
Activation failed. Timeout occurred.
Console side did not receive any
response from Agent\ThingWorx side
in predefined activation time delay.
Ensure network connection and retry.
Activation failed. Please contact your GE
HealthCare service personnel. (Error code:
400)
Bad request
Contact AVURI Device Connect
Support at
Activation failed. Please contact your GE
HealthCare service personnel. (Error code:
500)
Internal server error
Contact AVURI Device Connect
Support at
Activation failed. Please contact your GE
HealthCare service personnel. (Error code:
503)
Service unavailable
Retry
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......