GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 2 Site Preparations
2 - 15
2-3-6
Privacy and security
Since the LOGIQ systems are integrated into your IT-network, GE HealthCare wants to make sure that
you are aware of the proactive measures we are taking to secure the LOGIQ Totus.
For more information on privacy and security, “Privacy and Security Manual” (latest revision).
2-3-7
Environmental dangers
Commercial devices such as laser cameras, printers, VCRs and external monitors, usually exceed
allowable leakage current limits and, when plugged into separate AC outlets, are in violation of patient
safety standards. Suitable electrical isolation of such external AC outlets, or providing the device with
extra protective earth, will be required in order to meet ANSI/AAMI ES60601-1 and IEC60601-
1standards for electrical leakage.
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......