GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
10 - 28
Section 10-6 - Electrical Safety Tests
When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet
and the equipment under test is plugged into the receptacle on the panel of the meter. This places the
meter in the grounding conductor and the current flowing from the case to ground will be indicated in
any of the current ranges. The maximum allowable limit for enclosure leakage current is shown in
10-6-5-3
Data Sheet for Enclosure Leakage Current
The test passes when all readings measure less than the value shown in Table 10-9.
Record all data on the Ultrasound Inspection Certificate.
Table 10-14 Typical Data Sheet for Enclosure Leakage Current
Tester
Neutral
Tester Polarity
Switch
Protective Earth
Switch
Test 1
Wheel
Test 2
LCD
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
Closed
Normal
Closed
Closed
Normal
Open
Closed
Reversed
Closed
Closed
Reversed
Open
Open
Normal
Closed
Open
Reversed
Closed
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......