GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-63
8-3-11
Functional Check
8-3-11-1
Service Platform
1.) Click on
Utility > Service.
It will take about ten (10) seconds for activating.
2.) Make sure that
CAPS
is OFF (should be dimmed) for password entry performed later.
3.) The Service Login window for Service Platform will be shown on the monitor display.
Figure 8-187 Service
Figure 8-188 CAPS OFF
OFF
ON