GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
8-60
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-10-9
Regional Preset
8.) Touch
Utility > System
.
9.) Click
About
tab.
10.)Click
Additional About Information
.
11.)Check the region name of
Preset Region (Rquires reboot):
If it is None, skip this section.
12.)Click
Reload Factory Default
button.
13.)Click
OK
on confirmation dialog box.
This changes Preset Region (Requires reboot) to
None
.
NOTICE
For the system other than "NONE" selected as Regional Preset, resetting MUST be required.
Figure 8-182 Preset Region
Figure 8-183 Reload Factory Default