GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
8-52
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-10-6 Regional Setup (For the system with language other than English ONLY)
3.) Select the
Language tab
, press
Details
, under Installed Services select the
Russian or Greek
keyboard
, under Default input language select
Russian - Russian or Greek - Greek
, press
Apply
,
Press
OK
.
4.) Select the
Advanced tab
, then select
Russian or Greek
in the Language for non-Unicode
programs pull-down menu. Press
Apply
. Answer
Yes
to use files already loaded on the hard disk,
then answer
No
to not reboot the system yet, press
OK
. Press
Save
and Exit the Utility screen.
5.) Reboot the system. When your system restarts, the system appears in the selected language.
6.) To switch between the English and Russian (or Greek) keyboard, press
Alt
+
Shift
to change the
keyboard to the Russian or Greek keyboard.
7.) Apply the changes by pressing
Apply
. Press
OK
TWICE
.
Figure 8-173 Advanced Tab
NOTICE
To have the settings take effect, you MUST turn off the system and turn it back on.