GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-13
8-3-6
Base System Installation (Ghost) and Application Load
This section describes actual system installation and application loading procedures.
NOTE:
It takes approx. 25 minutes to complete software loading. It is longer than software loading to
the BEP3 installed system. Slow response or delayed display is normal during software loading.
8-3-7
Install Base System
1.) Power OFF the scanner. Wait 5 or more seconds for PC protection.
2.) Then power ON the scanner, and open the DVD drive and set the Ghost DVD.
3.) DVD drive reads Ghost DVD automatically and following screen appears, press the
Enter
key.
Figure 8-120 Loading Software