GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-49
8-3-10-6 Regional Setup (For the system with language other than English ONLY)
9.) Select the
Advanced tab
, then select
Japanese
in the Language for non-Unicode programs pull-
down menu. Press
Apply
. Answer
Yes
to use files already loaded on the hard disk, then answer
No
to not reboot the system yet, press
OK
. Press
Save
and Exit the Utility screen.
10.)Reboot the system.
11.)In
Utility--> System--> General
, press
Regional Options
. Select
Language Tab
and press
Details
.
Remove English
under Installed Services and press
Apply
and
OK
.
12.)Select
Utility--> System--> About-->Additional Info menu
.
13.)Select
Automatic Setup for Japanese Language
button. Then IME configuration shall be
automatically done.
This configuration shall setup the keyboard assignment (remove
半角かな
, etc) in IME.
14.)Reboot the system.
NOTICE
To have the settings take effect, you MUST turn off the system and turn it back on.