Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-9
8-3-5
Pre-installation Procedures
The following section describes activities necessary before system and application installation.
8-3-5-1
Disabling USB Devices
1.) Disconnect all USB peripherals connected to the scanner.
8-3-5-2
Save Customer Data / Setting
1.) Save the following user data.
-
Patient image data (Refer to operation manual.)
2.) Confirm if the customer has any Not-Finalized-Media (CD/DVD-R), used on the current version. All
Not-Finalized-Media must be finalized before the system is upgraded to BT08.
8-3-5-3
Operator Login Password
1.) The Operator Login Password will be deleted even if the BECOMP is not replaced. So, write down
the password if necessary.
CAUTION
Loading R7.7.x application software procedures requires basically C drives to be ghosted. In
this case, patient image data, preset parameters, and customer data will be kept. However , in
case full ghost will be required, you MUST back up before starting this procedure as follows.
For customer data safe, we recommend you to back up even for the C ghost only.
Figure 8-116 Operator Login Window
Table 8-62 Recording Operator Login Password
Parameters
Descriptions
Operator Login Password