Section 6-2 - Caster Brake/Swivel Function Adjustments
6-2-1-3 Caster lock/swivel does NOT lock even if the pedal is locked in position
2.) Rotate the right adjusting bolts
two turns
as shown. Perform this procedure at the left adjusting bolt.
3.) Tighten the left and right adjusting lock bolts.
4.) Verify that lock and brake is applied when the brake pedal is depressed. If it is not, repeat steps 1
to 3.
Figure 6-79 Rotate the bolts two turns