GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Chapter 8 -
Page 8-65
8.) Unplug the J14 connector, and unscrew 2 screws. Pull out the APS II Assy carefully.
Note: AC power input connector should be passed through the square hole on chassis bottom.
9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-25-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-104 Pull out the APS Assy
Table 8-26 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-25. Equipment
passed all required tests and is ready for use. “
APS II Assy PCB Function Validation