GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-72
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-29
CWD64 Assy
8-2-29-1
Tools
•
Common Phillips screwdrivers
8-2-29-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-29-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-48
.
2.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-63
3.) Remove the RFC Assy. Refer to the
8-2-32 "RFC Assy" on page 8-78
.
4.) Disconnects the CWD64 cable.
5.) Pull out the CWD64 Assy located in #3 from Nest box.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-29-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-111 CWD64 Assy
Table 8-30 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-29. Equipment
passed all required tests and is ready for use. “
CWD64 Assy Function Validation Procedure
Functional Checks (See Also Chapter 4)
CWD64 Assy
CWD64 cable