GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-40
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-8
Trackball Assy
8-2-8-1
Tools
•
Common Phillips screwdrivers
8-2-8-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-8-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
8-2-6 "Main Keyboard Assy" on page 8-37
2.) Unscrew 4 screws (1-4) and disconnect the cable.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-8-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-75 Screws and connector for Trackball
Table 8-9
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-8. Equipment
passed all required tests and is ready for use. “
4 screws
connector