Chapter 8 -
Page 8-39
8-2-7
Sub Keyboard Assy
8-2-7-1
Tools
•
Common Phillips screwdrivers
8-2-7-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-7-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
8-2-6 "Main Keyboard Assy" on page 8-37
2.) Unscrew 4 screws. 2 screws(1,3) are back of the speaker enclosures, and 2 screws(2,4) are bottom
of sub keyboard.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-7-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-74 4 Screw Sub Keyboard Assy
Table 8-8
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-7. Equipment
passed all required tests and is ready for use. “