![background image](http://html.mh-extra.com/html/ge/druck-adts-2-series/druck-adts-2-series_user-manual_132649045.webp)
33
Druck ADTS 2XX User Manual
K0451 Issue No. 1
SOUR:RATE
Command Syntax
SOURce:RATE<parameter> <aim>
Parameter Type
DISCRETE, REAL
Static Channels:
SOUR:RATE <parameter>,<Ps1 Rate Aim>,<Ps2 Rate Aim>
Pitot Channel:
SOUR:RATE <parameter>,<Pt Rate Aim>
Parameter Range:
ALT | CAS | MACH | PS | PT | QC
Function:
To command the rate of change of the selected parameter to a new aim
value.
Conditions:
Must be in control mode.
Query Syntax
SOURce:RATE? <parameter>
Parameter Type
DISCRETE
Parameter Range:
ALT | CAS | MACH | PS | PT | QC
Returned Static Channels:
<Ps1 Aim>,<Ps2 Aim>
Returned Pitot Channel
<Pt Aim>, *
Function:
To query last rate command.
Conditions:
None.