4.3 Maintenance Tasks
A
Check that all the equipment is present; record any deficiencies. Visually inspect the
external of the ADTS 2XX, and its associated equipment, for obvious signs of damage,
dirt, and the ingress of moisture. If necessary, use mild liquid detergent (item 2, Table
4-2) and a lint-free cloth (item 1, Table 4-2) to clean the external surfaces.
Inspect the pressure outlet ports and expansion ports for ingress of dirt and moisture,
clean if necessary with a lint-free cloth.
B
Visually inspect the pneumatic output connectors for damage. Inspect the small o-ring
on each pneumatic output connector for cuts and any signs of wear; replace as
necessary. Visually inspect pneumatic hoses, electrical cables for cuts, splits and
damage; replace as necessary.
Check, by entering the system set-up, service data (figure 4-1 below) the unit and pump
operational hours. At 1000 hours (filters) and 3000 hours (pump) return the unit to the
servicing depot for maintenance.
C
Before use, power-up the unit as detailed in Section 3. Check the date of the last
calibration and, if necessary, refer to the manufacturer. Record any error messages
and refer to Section 5.
D
Daily and before use, carry out the SST and leak check detailed in Section 5.
F
IGURE
4-1 S
ERVICE
D
ATA
Содержание Druck ADTS 2 Series
Страница 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Страница 2: ... General Electric Company All rights reserved ...
Страница 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Страница 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...