Druck ADTS 2XX User Manual
26
K0451 Issue No. 1
DK286 – Touch Screen Calibrator
This program enables the calibration of the touch screen. To execute this application, exit from the ADTS 2XX
program and double click on the Calibrate Touch icon
. The Stylus application, stored as part of the
system application, executes first. Select the calibration option and click the
Recalibrate
button. Follow the
on-screen instructions, after obtaining all the calibration points, touch the screen once and the Stylus
program returns to the previous screen. Close down the Stylus application, the screen shows the following:
This screen gives the opportunity to save the calibration by clicking the
Save Screen Calibration
button, or
restoring the default screen calibration stored by clicking the
Restore Default Calibration Data
button. The
system stores the calibration data.
On completion, click the
Exit
button.
DK287 – Display Backlight Brightness Setting Program
This program enables the user to adjust the brightness of the display backlight. To execute this program, exit
the ADTS205 program and double click on the icon Set BkLight Brightness
. The screen shows the
following:
The slide bar Brightness Level starts in the position of the current backlight brightness. Adjust the brightness
level by moving the slide bar left for darker and right for brighter. The backlight changes as the slide bar
moves. To save the setting, click the
Save Brightness Setting
button, this stores the current brightness
setting.
On completion, click the
Exit
button.
Содержание Druck ADTS 2 Series
Страница 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Страница 2: ... General Electric Company All rights reserved ...
Страница 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Страница 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...