5
Druck ADTS 2XX User Manual
K0451 Issue No. 1
2.4
Electrical Connection
WARNINGS:
1
VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50 VOLTS DC, IN
CERTAIN CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE
TAKEN WHEN WORKING ON LIVE, EXPOSED CONDUCTORS.
2
DO NOT DISCONNECT THE ADVANCED HAND TERMINAL WHEN
ENERGIZED IN THE HAZARDOUS AREA.
U
SE
IN
A
HAZARDOUS
AREA
S
PECIAL
C
ONDITIONS
APPLY
TO
THE
ATEX
CERTIFIED
A
DVANCED
H
AND
T
ERMINAL
REFER
TO
PAGE
V
.
Power Supply
Single phase, 90 to 132 VAC, 47 to 400Hz, 200 VA.
180 to 260 VAC, 47 to 66Hz, 300 VA.
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
A qualified technician (see page i) must carry out the following procedure.
C
AUTIONS
:
1
T
HE
SUPPLY
MUST
PROVIDE
CONNECTION
TO
A
PROTECTIVE
GROUND
TERMINAL
. T
HE
UNIT
MUST
,
AT
ALL
TIMES
,
BE
CONNECTED
TO
THE
SUPPLY
EARTH
(
GROUND
).
2
T
HE
POWER
SUPPLY
CABLE
AND
CONNECTOR
MUST
BE
CORRECTLY
RATED
FOR
THE
POWER
SUPPLY
.
• Make sure that the power supply is off before connecting the power cable.
Fuses
The two fuses, located in the holders and mounted on the front panel, protect the unit. The fuses
are connected in the live and neutral supply circuit and are rated at:
5A anti-surge HBC 250V
External earth/ground connection
z
An external earth (ground) cable may be connected to the stud on the front panel of the unit
providing integrity of the earth (ground) connection.
Advanced Hand Terminal Power Pack (Figure 2-3)
z
Four interchangeable adaptors, supplied with the power pack, can be used to connect to the local
power supply. To change the power pin adaptor, push the adaptor from the power pack body,
align the replacement adaptor and push into the power pack body.
E
UROPEAN
C
OLOUR
U.S.
C
OLOR
F
UNCTION
Brown
Black Live
Blue
White
Neutral
Green/Yellow
Green
Protective Earth
(Ground)
Содержание Druck ADTS 2 Series
Страница 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Страница 2: ... General Electric Company All rights reserved ...
Страница 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Страница 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...