Druck ADTS 2XX User Manual
K0451 Issue No. 1
ATEX C
ERTIFIED
A
DVANCED
H
AND
T
ERMINAL
CONDITIONS OF USE
The ATEX certified Advanced Hand Terminal can be used in zone 2
*
hazardous areas in accordance with the
ATEX certification document and schedule.
Marking detail:
Refer to Advanced Hand Terminal User Manual K0418 and the label on the Advanced Hand Terminal.
SPECIAL CONDITION OF USE
•
The power supplies must be isolated when connecting the advanced hand terminal in the hazardous
area.
•
The advanced hand terminal must not be disconnected when energized in the hazardous area.
•
The advanced hand terminal is a non-serviceable component. If the advanced hand terminal
becomes unserviceable it can only be replaced by another ATEX compliant hand terminal.
Note:
The advanced hand terminal must only be used with the cable assembly supplied and marked “DO NOT
SEPARATE WHILST ENERGISED IN HAZARDOUS AREA”
*
Zone 2 hazardous area definition, see page 53.