Chapter 3. Logging Data
58
DigitalFlow™ GF868 Programming Manual (2-Channel)
3.5 Creating an ERROR Log
Use the
ERROR
submenu to create a new error log and to select the logging parameters. An error log updates every 5
seconds (when the display updates), but only if a new error condition occurs. Error logs have a fixed length of 2 pages
and contain sixty records per page. Each record shows the time of the error, the measurement parameter values at that
time, and the error code message. The logged values of the chosen measurement parameters at the time of the error
condition provide valuable troubleshooting information.
Press the appropriate side of the
[SCREEN]
key to activate the desired pane of the display screen, and complete the
following steps:
1.
To access the
Log Menu
, press the
[LOG]
key.
a.
If you have activated the security feature (see Chapter 1,
Programming Site Data
, page 35 for details), enter the
assigned password and press
[ENT]
.
2.
Press
[F4]
=
ERROR
.
3.
Use the [
, [
and
[F1]-[F4]
keys to inspect a completed log or to change an active log, or enter a new
Name
(up
to 5 characters) and press
[ENT]
to create a new log file.
Note:
The option bar lists all error logs currently stored in memory. Recall that a flashing asterisk (*) appears at the
far right of the locator bar if the Model GF868 is currently logging data (see Chapter 3,
Operation
, of the
Startup Guide
).
4.
At the Log Message prompt, enter a brief description (up to 21 characters) of the log. Press
[ENT]
.
5.
Use the [
, [
and
[F1]-[F4]
keys to select the first parameter to be logged. See Table 17 for a list of the available
options.
IMPORTANT:
Only the first parameter logged may be graphically displayed (see Chapter 2,
Displaying Data
).
Table 17: Measurement Parameter Options
Option Bar Choice
Description
[F1]
=
VEL
Flow Velocity
[F2]
=
VOLUM
Volumetric Flow
[F3]
=
+TOTL
Forward Totalized Volume Flow
[F4]
=
-TOTL
Reverse Totalized Volume Flow
[
+ [F1]
=
MDOT
Mass Flow
[
+ [F2]
=
+MASS
Forward Totalized Mass Flow
[
+ [F3]
=
-MASS
Reverse Totalized Mass Flow
[
+ [F4]
=
DIAG
Diagnostic
Содержание DigitalFlow GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 60: ...Chapter 2 Displaying Data 50 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 95: ...DigitalFlow GF868 Programming Manual 2 Channel 85 Appendix A Menu Maps Appendix A Menu Maps ...
Страница 96: ...86 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 104: ...Appendix A Menu Maps DigitalFlow GF868 Programming Manual 2 Channel 94 ...
Страница 108: ...98 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 146: ...Warranty 136 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 150: ......
Страница 151: ......