Chapter 1. Programming Site Data
32
DigitalFlow™ GF868 Programming Manual (2-Channel)
1.7.4c Procedure Options
After completing the above steps, the meter returns to the
Global PROGRAM
prompt. Continue as follows:
•
To continue programming the meter, refer to the menu maps in Appendix A and navigate to the desired menu.
Then, proceed to the appropriate section of this manual for instructions.
•
To leave the
User Program
and retain the previous settings, press
[EXIT]
twice and then press
[F1] = NO
at the
SAVE
prompt. Any programming changes will be discarded and you will be returned to the data display.
•
To leave the
User Program
and return to measurement mode, press
[EXIT]
twice and then press
[F2] = YES
at the
SAVE
prompt. Your programming changes will be entered into the meter’s memory, and you will be returned to
the data display.
Note:
See the
SAVE
submenu section on page 37 of this chapter for more information on this function.
1.7.5 MODBUS Register Map
When equipped with the optional MODBUS output card, the GF868 flow transmitter can send flow data and diagnostic
information to a flow computer (or SCADA) serially, using a Gould-type RTU protocol. In this case, only the
MODBUS function command, 3 (read multiple registers), 6 (write multiple registers) is valid. The format for the data
exchange is as follows:
•
The
send
command (initiated by the host flow computer or controller) comes in the form:
[time delimiter]<Addr><3><First Register MSB>
<First Register LSB><Register Count MSB>
<Register Count LSB><CRC Low><CRC High>[time delimiter]
•
The response (initiated by the host flow computer or controller) comes in the form:
[time delimiter]<Addr><3><Byte count><Data.........>
<CRC Low><CRC High>[time delimiter]
The format for the returned data types is as follows:
•
Integer (16 bit Integer) <MSB><LSB>
1 Register - 16 bit integer
•
Integer (32 bit IntegerI) <MSB><LSB><LSB><LSB>
2 Register - 32 bit integer
•
Floating Point (FP) <EXP><MAN><MAN><MAN>
2 Registers - 32 bit IEEE floating point number
To request specific parameters from the GF868 via the MODBUS, the control system must enter the appropriate
register number. Only registers 1 through 90 are available for MODBUS communications, while registers 508 through
512 are used by the GF868 to store the MODBUS parameters. For details, see Table 8 on page 33.
Содержание DigitalFlow GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 60: ...Chapter 2 Displaying Data 50 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 95: ...DigitalFlow GF868 Programming Manual 2 Channel 85 Appendix A Menu Maps Appendix A Menu Maps ...
Страница 96: ...86 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 104: ...Appendix A Menu Maps DigitalFlow GF868 Programming Manual 2 Channel 94 ...
Страница 108: ...98 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 146: ...Warranty 136 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 150: ......
Страница 151: ......