Appendix C. Programming with PanaView
110
DigitalFlow™ GF868 Programming Manual (2-Channel)
C.5.6 Handling Site Files
To access site files, click on the
Site Files
button in the
Properties
window. The
Site File Operations
window (shown in
Figure 31:
Site File Operations
Window
C.5.6a Saving an Existing Site to the Meter
To save an existing site to the meter:
1.
Select the radio button for
Selected
and highlight an existing site in the left pane.
2.
Then click on the
Save Site to Meter
button. A screen opens similar to Figure 32.
Figure 32: Site File Selection
3.
Highlight the desired site and click
Open.
PanaView sends the site to the meter.