Chapter 1. Programming Site Data
8
DigitalFlow™ GF868 Programming Manual (2-Channel)
1.5 The User Program
Use the keypad, as described in Chapter 3,
Operation
, of the
Startup Guide
to navigate through the
User Program
. The
menu maps in Appendix A may be followed in sequence, or the [
and [
keys may be used to scroll through the
prompt screens. The [
key may be used to delete the last alphanumeric character that was entered from the keypad.
The following discussion assumes that the left screen pane is active. If the right screen pane is active, only the function
key designations change. That is, replace
[F1]-[F4]
with
[F5]-[F8]
. Also,
Channel 1
is used in all examples, but the
instructions apply equally to
Channel 2
.
Note:
Be sure to record all the programming data entered in this chapter in Appendix B,
Data Records
.
Programming of the
ACTIV
,
SYSTM
(
Ch1
and
GLOBL
) and
PIPE
submenus is necessary for operation of the Model GF868.
Failure to accurately enter all of the required information will result in unreliable flow rate data. Therefore, be sure to
complete at least the sections of this chapter pertaining to those three submenus.
Note:
Because it is so essential, instructions for programming
the
ACTIV
,
SYSTM
and
PIPE
submenus are also included
in the
Startup Guide
. If that programming has already been completed, skip the corresponding sections of this
chapter.
Except for the three submenus noted above, it is not necessary to program the Model GF868 flowmeter in any
particular order. Therefore, the sections of this chapter need not be completed in sequence. Proceed immediately to any
section of interest.
1.5.1 Entering Programming Mode
To access the
User Program
, press the
[PROG]
key on the keypad. The standard measurement mode screen will be
replaced by the following initial programming mode screen:
Note:
If the security feature is active, enter the password and press the
[ENT]
key to enter the
User Program
. See the
SECUR
submenu
section of this chapter on page 35 for more information on the security feature.
PROGRAM
Start
Press the [
and [
keys and one of the function keys
[F1]-[F4]
to select the
desired submenu. From this screen, press
[EXIT]
to return to measurement mode.
PROGRAM
Channel 1
CH1
CH2
GLOBL
SAVE
Содержание DigitalFlow GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 60: ...Chapter 2 Displaying Data 50 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 95: ...DigitalFlow GF868 Programming Manual 2 Channel 85 Appendix A Menu Maps Appendix A Menu Maps ...
Страница 96: ...86 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 104: ...Appendix A Menu Maps DigitalFlow GF868 Programming Manual 2 Channel 94 ...
Страница 108: ...98 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 146: ...Warranty 136 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 150: ......
Страница 151: ......