Chapter 2. Displaying Data
46
DigitalFlow™ GF868 Programming Manual (2-Channel)
2.5.3 Graphical Format (cont.)
In the typical graphical display shown, the first logged parameter
(VEL)
is shown to the left of the y-axis and there is a
message line
below the graph that initially indicates the starting date and time of the current page. Notice that most of
the locator bar has been replaced by a
status line
in standard video that shows the value, units and time of the record at
the current cursor location. However, there is still an inverse video
pointer
at the far right of this line to indicate that
additional option bar choices are available.
Note:
The date in the message line includes only the day and the month (not the year). The times in the status line and
the message line include only hours and minutes (not seconds).
Because the times displayed in the
LOG
graph are in whole minutes only, movement of the cursor may not result in a
visible change in the posted times. For example, if a
TIME INCREMENT
of 30 seconds has been programmed, two presses
of either cursor movement key will be required to change the displayed time by one minute.
2.6 Displaying the Transducer Signal
The
SIGNL
submenu permits the direct graphical display of several transducer signals. Specifically, the signals listed in
Table 13 may be viewed in this format.
Note:
The Measure type signals (Mup and Mdown) are available only if the
S/M
burst technique was selected in the
ACTIV
menu.
[SCREEN]
key to activate the desired pane of the display screen, then proceed as follows:
Note:
For this discussion, it is assumed that the left pane of the display screen is active. If the right pane is active, it
is only necessary to replace the
[F1]-[F4]
keys designations by the
[F5]-[F8]
keys.
1.
To access the
Display Menu
, press the
[DISP]
key.
2.
Press [
and
[F1]
=
SIGNL
.
3.
Press
[F1]
to display the
Skan
type signals or press
[F2]
to display the
Measure
type signals.
Table 13: Available Transducer Signals
Transducer Signal
Description
Skan Types
Sup
upstream Skan signal
Sdown
downstream Skan signal
Cup
upstream Skan correlation
Cdown
downstream Skan correlation
Measure Types
Mup
upstream Measure signal
Mdown
downstream Measure signal
Содержание DigitalFlow GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 60: ...Chapter 2 Displaying Data 50 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 95: ...DigitalFlow GF868 Programming Manual 2 Channel 85 Appendix A Menu Maps Appendix A Menu Maps ...
Страница 96: ...86 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 104: ...Appendix A Menu Maps DigitalFlow GF868 Programming Manual 2 Channel 94 ...
Страница 108: ...98 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 146: ...Warranty 136 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Страница 150: ......
Страница 151: ......