Error Codes and Screen Messages
2-7
March 2008
Log Active, END only
hit any key
When in the LOG menu, this message indicates that the log is still compiling
data. Only the End Time may be edited.
Log DONE,
to inspect hit any key
When in the LOG menu, this message indicates that the log is complete. Hit
any key to display the log.
Log hasn’t started! hit key
When in the LOG menu, this message indicates that the log has not yet
started.
OK, settling
The measurement has not stabilized. Wait before taking a reading.
Outside limits,
value rejected.
When in the CAL menu, this message indicates that the calibration of the
analog output is invalid. Hit any key to clear the message, and the DF868
will default to the last valid calibration. The message will also appear if there
is no ammeter connected to the analog output during calibration.
Overflow
The display value overflowed. Reduce the number of decimal digits or
change the units.
??P<L Enter L again.
The entered path length (P) is less than the axial dimension (L). Enter a new
value for L.
range is X.XX to X.XX
When in the
User Program
, this message indicates that the entered number
is unacceptable. Hit any key and enter a number within the allowable range.
Review calibration
An option card error indicating a programming failure or a loss of memory.
Contact the factory.
Review parameters
An option card error indicating a programming failure or a loss of memory.
Contact the factory.
Staring time must exceed
current time
This message appears when in the LOG menu STD option. Enter a start time
that is later than the current time.
Write error.
An option card error indicating a programming failure or a loss of memory.
Contact the factory.
Table 2-1: Screen Messages
Message
Meaning
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...