March 2008
B-4
Optional Enclosures
Rack Mount Option Card
Installation (cont.)
6.
Place the metal bracket over the option cards, making sure that all
installed option cards are inserted into the plastic card guides in
the bracket. The bracket must be oriented so that the six card
guides are directly above the six sockets on the printed circuit
board (
do not
install the bracket rotated 180° out of position).
Secure the metal bracket to the printed circuit board with the four
plastic snap rivets provided.
7.
After checking for any loose hardware that may have fallen into
the enclosure, reinstall the top panel on the meter and secure the
panel in place with the four screws previously removed.
The Model DF868 flowmeter may now be placed back into service.
Reconnect the main power and refer to Appendix C,
Optional
Enclosures
, of the
Startup Guide
for instructions on wiring the new
option card(s).
Note:
Be sure to record the option card installation in Appendix A,
Service Record
.
Содержание DigitalFlow DF968