Calibration
1-13
March 2008
Entering Pulse Number
5.
Enter the number of pulses desired (between 1 and 10,000) and
press the
[ENT]
key. That number of pulses will then be output at
the specified frequency.
6.
Repeat steps 3, 4 and 5 to test all four of the frequency/totalizer
outputs. If any of the outputs fails to pass the test, contact the
factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of
the following:
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to
the appropriate section.
•
To leave the
Calibration Menu,
press
[EXIT]
twice
.