Error Codes and Screen Messages
2-5
March 2008
E13: Over Range
Problem:
This error code message indicates that the present
measurement exceeds the capacity of the meter.
Cause:
A internal mathematical overflow has occurred in either
the volumetric or mass flow calculations.
Action:
Select larger measurement units or a shorter time interval
for the current measurement parameter. For example,
choose
KSCF/M
instead of
SCF/M
. See the
Startup Guide
for instructions.
E14: Totals Overflow
Problem:
The totalizers are unable to keep up with the total flow
signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value.
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...