Error Codes and Screen Messages
E8: Temp In Supply
Problem:
This message indicates a temperature input supply error.
Cause:
The temperature exceeds the specified limits for the
analog inputs option card.
Action:
Check the temperature transmitter and the connecting
cable. Refer to Chapter 1,
Calibration
, and recalibrate the
analog inputs on the option card (page 1-6).
E9: Temp In Return
Problem:
This message indicates a temperature input return error.
Cause:
The temperature exceeds the specified limits for the
analog inputs option card.
Action:
Check the pressure transmitter and the connecting cable.
Refer to Chapter 1,
Calibration
, and recalibrate the
analog inputs on the option card (page 1-6).
E10: Weak Signal
Problem:
This message indicates a weak signal.
Cause:
The magnitude of the two-phase signal is not large
enough or the flow rate in the pipe is out of range.
Action:
Check the pipe and setup parameters and the transducer
connections. The process may not include enough two-
phase content — switch to transit-time mode.
E11: Poor Signal
Problem:
This message indicates a poor signal.
Cause:
The shape of the processed signal is not good enough to
formulate a flow measurement.
Action:
Check the pipe and setup parameters and the transducer
connections. The process may not include enough two-
phase content — switch to transit-time mode.
E12: Incoherent Signal
Problem:
This message indicates an incoherent signal.
Cause:
The flow measurements, as determined by the two
Repetition Periods, are too dissimilar.
Action:
Check the pipe and setup parameters and the transducer
connections. The process may not include enough two-
phase content — switch to transit-time mode.
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...