March 2008
1-6
Calibration
Calibrating the Analog
Inputs
Analog inputs may be added to the Model DF868 flowmeter by
installing an
Analog Inputs Option Card
in one (or more) of the six
expansion slots. The option card contains two analog inputs, which
are designated as A and B. Each of the inputs must be calibrated at
both the zero-point and full-scale values. After calibrating the inputs,
their linearity should be tested.
Calibration of the analog inputs requires the use of a calibrated
current source. If an independent calibrated current source is not
available, one of the
Slot 0
analog outputs may be used for the
calibration. During the analog input calibration, the
Slot 0
analog
output will supply the low reference, high reference, 4 mA and 20
mA signals at the appropriate times.
Note:
If a
Slot 0
analog output will be used to calibrate the analog
inputs, make sure that the
Slot 0
analog output calibration
procedure has already been completed.
Connect the
Slot X
analog output(s) (or an independent calibrated
current source) to the analog input(s) on the option card, as shown in
Figure 1-3 below. Refer to the menu map in Figure 1-6 on page 1-15
as a guide.
Figure 1-3: Analog Input Calibration Connections
Analog Input Option Card
A
INLO
A
RTN
A
+24V
A
INHI
8
1
B
INLO
B
RTN
B
+24
B
INHI
B
RTN
B
SIG
A
RTN
A
SIG
Terminal Block I/O (Slot 0)
4
1
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...