Calibration
1-9
March 2008
Calibrating the RTD
Inputs
Calibrating an RTD option card involves a slightly different
procedure than for other analog input cards. However, you access the
card in the same manner.
Accessing the Calibration
Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include
a slot listing for each installed option card.)
3.
Press
[F1]
to select Input A, or
[F2]
to select Input B.
4.
Press
[F1]-[F2]
to select the reference point to be calibrated.
Do one of the following:
•
If you selected
[F1]
(LOW),
proceed to step 5.
•
If you selected
[F2] (HIGH),
proceed to step 8.
LOW Option = [F1]
5.
Enter the low reference value and press the
[ENT]
key.
6.
Press
[F1]
to
STORE
the low reference value or
[F2]
to
ABORT
the
calibration. In either case, the screen returns to the
ANALOG INPUT
prompt.
7.
Press
[F2]
to proceed to the next section.
HIGH Option = [F2]
8.
Enter the high reference value and press the
[ENT]
key.
9.
Press
[F1]
to
STORE
the current high reference value or press
[F2]
to
ABORT
the entry. In either case, the
ANALOG INPUT
prompt will
reappear.
10.
Proceed to
Procedure Options
below.
Procedure Options
You have completed calibration of the RTD input. Do one of the
following:
•
To calibrate another input, press
[EXIT].
•
To calibrate additional slots, press
[EXIT]
and then
[F4] (EXIT)
and
proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...