Calibration
1-7
March 2008
Calibrating the Analog
Inputs (cont.)
For this discussion, assume that the option card has been installed in
Slot x
.
Note:
The zero point of the analog input may be set for either 0 mA
or 4 mA. However, the calibration procedure always uses the 4
mA point, as the meter will extrapolate this value to obtain the
0 mA point.
Accessing the Calibration
Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include
a slot listing for each installed option card.)
3.
Press
[F1]
to select the desired input.
The procedure for calibrating
Input A
is identical to that for calibrating
Input B
. However, when calibrating
Input B
, be sure a calibrated
current source is connected to the appropriate pins on the terminal
block. See Figure 1-3 on page 1-6 for the correct pin numbers.
4.
Press
[F1]-[F4]
to select the reference point to be calibrated.
Do one of the following:
•
If you selected
[F1]
(aLOW),
proceed to
aLOW Option
on the next
page.
•
If you selected
[F2]
(aHIGH)
,
proceed to
aHIGH Option
on the next
page.
•
If you selected
[F3]
(4 mA),
proceed to
4 mA Option
on the next
page.
•
If you selected
[F4]
(20 mA)
,
proceed to
20 mA Option
on the next
page.
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...