Access to power switch and power cord. Do not block the rear and side panel of
the instrument. The Power switch must always be easy to access. The power cord
must always be easy to disconnect.
•
Installing the computer. The computer should be installed and used according to
the instructions provided by the manufacturer of the computer.
Notices
NOTICE
To avoid damage to ÄKTAavant or other equipment when installing
or moving the instrument, follow the instructions below.
•
Vents on the ÄKTAavant instrument. To ensure adequate ventilation, keep papers
and other objects away from the vents of the instrument.
•
Disconnect power. To prevent equipment damage, always switch off power to the
ÄKTAavant instrument before an instrument module is removed or installed, or a
cable is connected or disconnected.
•
Misuse of UniNet-9 connectors. The UniNet-9 connectors at the rear panel should
not be mistaken for Firewire connectors. Do not connect any external equipment to
the UniNet-9 connectors. Do not disconnect or move the UniNet-9 bus cable.
System operation
Warnings
WARNING
To avoid personal injury when operating the ÄKTAavant system,
follow the instructions below.
•
Rotating the instrument. Make sure that there is always at least 20 cm of free space
around the ÄKTAavant instrument to allow rotation on the swivel foot. When rotating
the system, take care not to stretch or squeeze tubing or cables. A disconnected
cable may cause power interruption or network interruption. Stretched tubes may
cause bottles to fall, resulting in liquid spillage and shattered glass. Squeezed tubing
may cause increase in pressure, or block liquid flow. To avoid the risk of turning over
bottles, always place bottles on the buffer tray, and close the doors before rotating
the instrument.
16
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
2 Safety instructions
2.1 Safety precautions