7.3 Prime buffer inlets and purge System pumps
Introduction
Before usage of the System pumps, it is important to:
1
Prime the inlets (fill the buffer inlets with liquid).
2
Purge the System pumps (remove air from the pumps).
Prime buffer inlets
Follow the instruction below to prime the buffer inlets. Each buffer inlet that is to be used
during the method run has to be primed.
Action
Step
Make sure that all buffer inlet tubing that is to be used during the method
run is immersed in the correct buffers.
1
Make sure that the waste tube connected to Injection valve port W1 is
immersed in a waste vessel.
2
Open the System Control module in UNICORN, and select Manual:Execute
Manual Instructions.
3
Result: The Manual instructions dialog opens.
78
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
7 Prepare the system for a run
7.3 Prime buffer inlets and purge System pumps
Содержание AKTA avant
Страница 1: ...GE Healthcare Getting Started with KTAavant and UNICORN 6...
Страница 2: ......
Страница 133: ......