Action
Step
Connect a 25-30 ml syringe to the left purge valve of the selected system
pump. Make sure that the syringe fits tightly into the purge connector.
2
Open the purge valve by turning it counterclockwise about 3/4 of a turn.
Draw 5-10 ml of liquid slowly into the syringe with a rate of about 1 ml/s.
3
Close the purge valve by turning it clockwise. Disconnect the syringe and
discard its contents.
4
Connect the syringe to the purge valve on the right pump head of the
selected system pump, and repeat steps 3-4.
5
82
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
7 Prepare the system for a run
7.3 Prime buffer inlets and purge System pumps