Column cleaning
After a method run is completed, perform the following:
•
Clean the column with one or several cleaning solution(s) using Column CIP.
Column storage
If the column is not going to be used for a couple of days or longer, also perform the
following:
•
Fill the column with storage solution (e.g., 20% ethanol) using Column CIP.
pH electrode storage
Follow the instruction below to fill the pH flow cell with storage solution. The calibration
function is used to switch the pH valve. However, no calibration is performed.
Action
Step
Open the System Control module and select System:Calibrate.
1
Result: The Calibration dialog opens.
In the Calibration dialog, select pH from the Monitor to calibrate drop-down
list.
2
Press
.
3
Result: The pH valve switches to the calibration position.
112
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
8 Run a method
8.5 After run procedures