GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 19
Direction 18000
6-9A-15
9A-7 Radial Alignment
If this is a detector installation, perform Radial alignment. Otherwise, skip this section and go directly to ISO
center. Radial alignment need only be done for detector installations or if the detector has been removed from
the gantry for any reason.
NOTE:
Verify that the Tube Unit is ‘‘cold”; i.e., that it has had no more than 25 kilojoules exposure
[KV
•
mA
•
Sec/1000] within the last 30 minutes. Wait a minimum of 90 minutes from the
completion of a generator cal or other high technic exposures. Each exposure in the
procedure adds 12.8kj of heat to the tube, exposures should be made no faster than one every
5 minutes. Tube warm-up is NOT recommended prior to this procedure.
NOTE:
A rough ISO center adjustment must be made prior to radial alignment. Use the ISO center
procedure (section 9A-8) and move the tube so that it is within .005” (0.10 channels) of the
correct position. Don’t bother trying to tweak it in perfectly at this point. When ISO is roughly
set, start with step 1 of Radial alignment.
Requirement: The initial starting position of the cam must be in the center of its range of motion
(Installation instructions, step 11-F).
1) Using option 3 in DC, move the tube to 0. Do not leave the DC program for the remainder of the scanning
portion of this procedure (i.e. do not exit DC until after step 4).
2) Turn off the scan breaker. Open the front gantry cover.
3) Install the rotational alignment tools on the detector. The pin must project into the X-ray beam.
See figure 6-9A-11.
4) Using default scan #9, perform a stationary scan of the pin. Record the exam number.
5) Exit DC and enter AN. After entering the exam number (if necessary), [RETURN] to get to the Scan Options
menu.
Содержание Advantium CT9800
Страница 2: ......
Страница 10: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 16: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 i A 6 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 25: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 4 1 5B THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 30: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 1 10 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 44: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 2 14 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 46: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 2 ILLUSTRATION 4 3 1 CC1 POWER SWITCHES ...
Страница 49: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 5 ILLUSTRATION 4 3 3 ...
Страница 53: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 9 ILLUSTRATION 4 3 8 ARGUS CABINET BREAKERS ...
Страница 95: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 24 Direction 18000 4 3 49B This page left intentionally blank ...
Страница 116: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 14 Direction 18000 4 3 70 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 119: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 9 3 ILLUSTRATION 4 9 1 LASER WARNING AND PRECAUTIONS ...
Страница 122: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 9 6 ...
Страница 138: ...Page Left Intentionally Blank ...
Страница 142: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 19 Direction 18000 4 10 18 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 144: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 4 10 20 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 150: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 4 10 26 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 174: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 20 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 176: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 22 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 178: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 190: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 4 13 8 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 196: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 1 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 198: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 2 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 216: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 3 18 3 4 Disk System Adjustments Continued ...
Страница 222: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 4 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 224: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 227: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 3 ILLUSTRATION 6 9 2 ...
Страница 234: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 10 ILLUSTRATION 6 9 8 ...
Страница 237: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 13 ILLUSTRATION 6 9 11 XENON DETECTOR BEAM ON WINDOW ...
Страница 240: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 16 ILLUSTRATION 6 9 12 ...
Страница 246: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 22 ILLUSTRATION 6 9 14 CBF ALIGNMENT ...
Страница 249: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 25 ILLUSTRATION 6 9 15 SYSTEM SAG ...
Страница 254: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9A 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 272: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 6 9A 14D CT GUIDE FOR TROUBLESHOOTING LOW SIGNAL OUTPUT ...
Страница 280: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9A 22 ILLUSTRATION 6 9A 14 CBF ALIGNMENT ...
Страница 283: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 6 9A 25 ILLUSTRATION 6 9A 15 SYSTEM SAG ...
Страница 290: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 6 ILLUSTRATION 6 10 2 LASER LIGHT ALIGNMENT TOOL ...
Страница 302: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 18 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 306: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 4 ILLUSTRATION 6 11 1 DISPLAYED IMAGE OF HEADHOLDER ...
Страница 308: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 6 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 311: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 3 ILLUSTRATION 6 A 2 S140 BMC ERCC SWITCHES JUMPERS ...
Страница 312: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 314: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 ii THIS PAE LEFT BLANK INTENTIONALLY ...
Страница 315: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 1 1 ...
Страница 316: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 1 2 ...
Страница 317: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 1 3 ...
Страница 318: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 7 1 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 319: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 1 ...
Страница 320: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 2 ...
Страница 321: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 3 ...
Страница 322: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 4 ...
Страница 323: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 5 ...
Страница 324: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 6 ...
Страница 325: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 7 ...
Страница 326: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 8 ...
Страница 327: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 9 ...
Страница 328: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 10 ...
Страница 329: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 11 ...
Страница 330: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 12 ...
Страница 331: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 13 ...
Страница 332: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 2 14 ...
Страница 333: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14A ...
Страница 334: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 335: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 15 ...
Страница 336: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 16 ...
Страница 337: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 17 ...
Страница 338: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 18 ...
Страница 339: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 19 ...
Страница 340: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 20 ...
Страница 341: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 21 ...
Страница 342: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 22 ...
Страница 343: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 23 ...
Страница 344: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 24 ...
Страница 345: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 25 ...
Страница 346: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 26 ...
Страница 347: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 27 ...
Страница 348: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 28 37 46 218868G1 ...
Страница 349: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 29 ...
Страница 350: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 30 ...
Страница 351: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 31 ...
Страница 352: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 32 ...
Страница 353: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 33 ...
Страница 354: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 34 ...
Страница 355: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 7 2 35 ...
Страница 356: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 36 ...
Страница 357: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37 ...
Страница 358: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37A ...
Страница 359: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 360: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 38 ...
Страница 361: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 39 ...
Страница 362: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 40 ...
Страница 363: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 41 ...
Страница 364: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 7 2 42 ...
Страница 365: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 43 ...
Страница 366: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 44 ...
Страница 367: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 45 ...
Страница 368: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 46 ...
Страница 369: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 47 ...
Страница 370: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47A ...
Страница 371: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 372: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48 ...
Страница 373: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48A ...
Страница 374: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 375: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 49 ...
Страница 376: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 50 ...
Страница 377: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 51 ...
Страница 378: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 52 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 379: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 1 ...
Страница 380: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 381: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 1 ...
Страница 382: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 2 ...
Страница 383: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 4 3 ...
Страница 384: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 4 4 ...
Страница 385: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 1 OVERTEMPERATURE SENSOR SCHEMATIC COMPUTER CABINET ...
Страница 386: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 2 OVERTEMPERATURE SENSOR UNIT COMPUTER CABINET ...
Страница 387: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 1 REMOTE CONTROL STATION POWER ...
Страница 388: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 2 REMOTE CONTROL STATION POWER ...
Страница 390: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 402: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 1 12 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 411: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 5 ...
Страница 420: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 16 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 428: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 4 8 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 462: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 472: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 34 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 478: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 8 7 6 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 483: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 8 8 5 ILLUSTRATION A POSITIONING THE DETECTOR ...
Страница 487: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 3 ...
Страница 489: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 5 ...
Страница 490: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 6 ...
Страница 491: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 7 ...
Страница 492: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 8 ...
Страница 493: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 9 ...
Страница 494: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 10 ...
Страница 495: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 11 ...
Страница 496: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 12 ...
Страница 497: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 13 ...
Страница 498: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 14 ...
Страница 499: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 15 ...
Страница 500: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 16 ...
Страница 501: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 17 ...
Страница 502: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 18 ...
Страница 503: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 19 ...
Страница 504: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 20 ...
Страница 506: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 22 ...
Страница 507: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 23 ...
Страница 508: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 24 ...
Страница 509: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 25 ...
Страница 510: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 26 ...