GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 6
Direction 18000
6-10-10
Proceed as follows to align the internal axial lights:
1.
Turn on the alignment lights via the gantry test switch.
2.
Verify that the collimator blades are closed.
3.
Back off the “D” locking screw. (Illustration 6-10-5)
4.
If necessary, adjust screws “A”, “B”, and “C” together in the same direction to position the laser light mirror
into the beam.
5.
Adjust screw “A” to move the beam arc vertically so that the beam fans across the gantry opening. Adjust
screw “A” until the line is visible on the collimator below.
6.
Alternately adjust screws “B” and “C” to move the beam horizontally (i.e., fan beam moves in and out of the
gantry) until it is aligned with the notched portion of the teflon end on the laser alignment tool (Ill. 6-10-2). If
this is done, the beam should align with the closed blades of the collimator aperture*.
* NOTE: Plane of Rotation must be done prior to the alignment light adjustment ,otherwise, the beam
may not align with the closed blades of the collimator aperture. Also, if the beam is not
parallel to the scan plane (i.e., scewed across collimator aperture) go to next step.
7.
If the beam is not parallel to the scan plane (i.e., scewed), adjust screws “E” and “F” to pivot the beam about
the vertical axis, until beam aligns with the closed blades on the collimator aperture. Beam still should lie
within the notch of the teflon end.
8.
Lock screw “D” carefully.
9.
Go to Sagittal Alignment (sec. 10-6).
Содержание Advantium CT9800
Страница 2: ......
Страница 10: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 16: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 i A 6 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 25: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 4 1 5B THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 30: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 1 10 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 44: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 2 14 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 46: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 2 ILLUSTRATION 4 3 1 CC1 POWER SWITCHES ...
Страница 49: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 5 ILLUSTRATION 4 3 3 ...
Страница 53: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 9 ILLUSTRATION 4 3 8 ARGUS CABINET BREAKERS ...
Страница 95: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 24 Direction 18000 4 3 49B This page left intentionally blank ...
Страница 116: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 14 Direction 18000 4 3 70 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 119: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 9 3 ILLUSTRATION 4 9 1 LASER WARNING AND PRECAUTIONS ...
Страница 122: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 9 6 ...
Страница 138: ...Page Left Intentionally Blank ...
Страница 142: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 19 Direction 18000 4 10 18 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 144: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 4 10 20 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 150: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 4 10 26 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 174: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 20 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 176: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 22 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 178: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 190: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 4 13 8 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 196: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 1 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 198: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 2 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 216: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 3 18 3 4 Disk System Adjustments Continued ...
Страница 222: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 4 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 224: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 227: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 3 ILLUSTRATION 6 9 2 ...
Страница 234: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 10 ILLUSTRATION 6 9 8 ...
Страница 237: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 13 ILLUSTRATION 6 9 11 XENON DETECTOR BEAM ON WINDOW ...
Страница 240: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 16 ILLUSTRATION 6 9 12 ...
Страница 246: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 22 ILLUSTRATION 6 9 14 CBF ALIGNMENT ...
Страница 249: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 25 ILLUSTRATION 6 9 15 SYSTEM SAG ...
Страница 254: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9A 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 272: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 6 9A 14D CT GUIDE FOR TROUBLESHOOTING LOW SIGNAL OUTPUT ...
Страница 280: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9A 22 ILLUSTRATION 6 9A 14 CBF ALIGNMENT ...
Страница 283: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 6 9A 25 ILLUSTRATION 6 9A 15 SYSTEM SAG ...
Страница 290: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 6 ILLUSTRATION 6 10 2 LASER LIGHT ALIGNMENT TOOL ...
Страница 302: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 18 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 306: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 4 ILLUSTRATION 6 11 1 DISPLAYED IMAGE OF HEADHOLDER ...
Страница 308: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 6 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 311: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 3 ILLUSTRATION 6 A 2 S140 BMC ERCC SWITCHES JUMPERS ...
Страница 312: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 314: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 ii THIS PAE LEFT BLANK INTENTIONALLY ...
Страница 315: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 1 1 ...
Страница 316: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 1 2 ...
Страница 317: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 1 3 ...
Страница 318: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 7 1 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 319: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 1 ...
Страница 320: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 2 ...
Страница 321: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 3 ...
Страница 322: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 4 ...
Страница 323: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 5 ...
Страница 324: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 6 ...
Страница 325: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 7 ...
Страница 326: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 8 ...
Страница 327: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 9 ...
Страница 328: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 10 ...
Страница 329: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 11 ...
Страница 330: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 12 ...
Страница 331: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 13 ...
Страница 332: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 2 14 ...
Страница 333: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14A ...
Страница 334: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 335: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 15 ...
Страница 336: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 16 ...
Страница 337: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 17 ...
Страница 338: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 18 ...
Страница 339: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 19 ...
Страница 340: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 20 ...
Страница 341: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 21 ...
Страница 342: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 22 ...
Страница 343: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 23 ...
Страница 344: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 24 ...
Страница 345: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 25 ...
Страница 346: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 26 ...
Страница 347: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 27 ...
Страница 348: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 28 37 46 218868G1 ...
Страница 349: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 29 ...
Страница 350: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 30 ...
Страница 351: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 31 ...
Страница 352: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 32 ...
Страница 353: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 33 ...
Страница 354: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 34 ...
Страница 355: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 7 2 35 ...
Страница 356: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 36 ...
Страница 357: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37 ...
Страница 358: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37A ...
Страница 359: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 360: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 38 ...
Страница 361: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 39 ...
Страница 362: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 40 ...
Страница 363: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 41 ...
Страница 364: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 7 2 42 ...
Страница 365: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 43 ...
Страница 366: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 44 ...
Страница 367: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 45 ...
Страница 368: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 46 ...
Страница 369: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 47 ...
Страница 370: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47A ...
Страница 371: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 372: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48 ...
Страница 373: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48A ...
Страница 374: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 375: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 49 ...
Страница 376: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 50 ...
Страница 377: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 51 ...
Страница 378: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 52 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 379: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 1 ...
Страница 380: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 381: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 1 ...
Страница 382: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 2 ...
Страница 383: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 4 3 ...
Страница 384: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 4 4 ...
Страница 385: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 1 OVERTEMPERATURE SENSOR SCHEMATIC COMPUTER CABINET ...
Страница 386: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 2 OVERTEMPERATURE SENSOR UNIT COMPUTER CABINET ...
Страница 387: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 1 REMOTE CONTROL STATION POWER ...
Страница 388: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 2 REMOTE CONTROL STATION POWER ...
Страница 390: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 402: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 1 12 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 411: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 5 ...
Страница 420: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 16 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 428: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 4 8 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 462: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 472: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 34 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 478: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 8 7 6 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 483: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 8 8 5 ILLUSTRATION A POSITIONING THE DETECTOR ...
Страница 487: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 3 ...
Страница 489: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 5 ...
Страница 490: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 6 ...
Страница 491: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 7 ...
Страница 492: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 8 ...
Страница 493: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 9 ...
Страница 494: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 10 ...
Страница 495: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 11 ...
Страница 496: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 12 ...
Страница 497: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 13 ...
Страница 498: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 14 ...
Страница 499: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 15 ...
Страница 500: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 16 ...
Страница 501: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 17 ...
Страница 502: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 18 ...
Страница 503: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 19 ...
Страница 504: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 20 ...
Страница 506: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 22 ...
Страница 507: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 23 ...
Страница 508: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 24 ...
Страница 509: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 25 ...
Страница 510: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 26 ...