GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 12
Direction 18000
6-9-9
The software program “Detector Alignment” with the three V block tools (46-229269G1) has been discontinued.
The detector is now aligned using polaroid film.
To adjust the detector Z position, loosen the 1/4-20 Hex socket cap screw, then loosen the .75” Hex lock nut.
Now turn the .560 Hex spacer the direction and number of flats indicated. Relock the .75” Hex nut to 20-30 ft.lbs.
and finally tighten the 1/4-20 Hex socket head cap screw to 13 + 1.3 ft.lbs. The adjustment is made to each of
the three mounting supports.
When a detector is replaced, the initial adjustment of Z position should be for a spacing of 0.190” (4.3mm)
between the rear edge of the detector box and the mounting surface in the gantry as shown in Section E-E of
Illustration 6-9-7.
Channel 1 of the detector is used for a Q Cal reference signal. This channel has an adjustable wedge masking
about 50% of the channel as shown in view B of Illustration 6-9-7. The channel intensity and wedge position are
checked in Section 9-12.
To shift the wedge, loosen the locknut 1-1/2 turns while keeping pressure on the spring washer. Move the wedge
by turning the eccentric adjustment. After adjustment, tighten the locknut.
The wedge position should be properly set even on a replacement detector. If it has been loosened and an
adjustment is required, preposition the wedge away from the center of the detector so it partially masks only
channel 1. The wedge will then be moved in while monitoring the intensity of channel 1, relative to channel 2.
9-3 Procedure Outline and Loading Software
Illustration 6-9-8 is a protocol flow diagram of the X-Ray alignment and functional check procedures for a Xenon
detector system.
1.
For a system functional check, perform all of the steps listed in Illustration 6-9-8.
2.
For a previously aligned system, one or more of the steps may be omitted on checkout following replacement
of the tube, collimator or detector. This assumes that the other two components are not altered.
3.
Default scans have been set-up in Data Collection for most X-Ray alignment scans. Refer to Direction
18008; RDCS; Chapter 4, Section 1-2. The default scans have been listed in Illustration 6-9-8.
Содержание Advantium CT9800
Страница 2: ......
Страница 10: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 16: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM MANUAL Rev 1 Direction 18000 i A 6 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 25: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 4 1 5B THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 30: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 1 10 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 44: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 2 14 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 46: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 2 ILLUSTRATION 4 3 1 CC1 POWER SWITCHES ...
Страница 49: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 5 ILLUSTRATION 4 3 3 ...
Страница 53: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 4 3 9 ILLUSTRATION 4 3 8 ARGUS CABINET BREAKERS ...
Страница 95: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 24 Direction 18000 4 3 49B This page left intentionally blank ...
Страница 116: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 14 Direction 18000 4 3 70 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 119: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 9 3 ILLUSTRATION 4 9 1 LASER WARNING AND PRECAUTIONS ...
Страница 122: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 4 9 6 ...
Страница 138: ...Page Left Intentionally Blank ...
Страница 142: ...GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM Rev 19 Direction 18000 4 10 18 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 144: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 4 10 20 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 150: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 4 10 26 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 174: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 20 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 176: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 22 THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 178: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 4 12 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 190: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 4 13 8 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 196: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 1 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 198: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 2 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 216: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 3 18 3 4 Disk System Adjustments Continued ...
Страница 222: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 4 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 224: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 227: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 3 ILLUSTRATION 6 9 2 ...
Страница 234: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 10 ILLUSTRATION 6 9 8 ...
Страница 237: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 13 ILLUSTRATION 6 9 11 XENON DETECTOR BEAM ON WINDOW ...
Страница 240: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 16 ILLUSTRATION 6 9 12 ...
Страница 246: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 22 ILLUSTRATION 6 9 14 CBF ALIGNMENT ...
Страница 249: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9 25 ILLUSTRATION 6 9 15 SYSTEM SAG ...
Страница 254: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 20 Direction 18000 6 9A 0A THIS PAGE LEFT INTENTIONALLY BLANK ...
Страница 272: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 6 9A 14D CT GUIDE FOR TROUBLESHOOTING LOW SIGNAL OUTPUT ...
Страница 280: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 12 Direction 18000 6 9A 22 ILLUSTRATION 6 9A 14 CBF ALIGNMENT ...
Страница 283: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 6 9A 25 ILLUSTRATION 6 9A 15 SYSTEM SAG ...
Страница 290: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 6 ILLUSTRATION 6 10 2 LASER LIGHT ALIGNMENT TOOL ...
Страница 302: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 10 18 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 306: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 4 ILLUSTRATION 6 11 1 DISPLAYED IMAGE OF HEADHOLDER ...
Страница 308: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 6 Direction 18000 6 11 6 THIS PAGE IS LEFT INTENTIONALLY BLANK ...
Страница 311: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 3 ILLUSTRATION 6 A 2 S140 BMC ERCC SWITCHES JUMPERS ...
Страница 312: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 6 A 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 314: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 ii THIS PAE LEFT BLANK INTENTIONALLY ...
Страница 315: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 1 1 ...
Страница 316: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 1 2 ...
Страница 317: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 1 3 ...
Страница 318: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 13 Direction 18000 7 1 4 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 319: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 1 ...
Страница 320: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 2 ...
Страница 321: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 3 ...
Страница 322: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 4 ...
Страница 323: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 5 ...
Страница 324: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 6 ...
Страница 325: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 7 ...
Страница 326: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 8 ...
Страница 327: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 9 ...
Страница 328: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 10 ...
Страница 329: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 11 ...
Страница 330: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 12 ...
Страница 331: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 13 ...
Страница 332: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 2 14 ...
Страница 333: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14A ...
Страница 334: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 14B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 335: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 15 ...
Страница 336: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 16 ...
Страница 337: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 7 2 17 ...
Страница 338: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 18 ...
Страница 339: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 19 ...
Страница 340: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 20 ...
Страница 341: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 9 Direction 18000 7 2 21 ...
Страница 342: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 22 ...
Страница 343: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 23 ...
Страница 344: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 24 ...
Страница 345: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 25 ...
Страница 346: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 26 ...
Страница 347: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 27 ...
Страница 348: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 28 37 46 218868G1 ...
Страница 349: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 29 ...
Страница 350: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 30 ...
Страница 351: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 31 ...
Страница 352: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 32 ...
Страница 353: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 33 ...
Страница 354: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 34 ...
Страница 355: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 10 Direction 18000 7 2 35 ...
Страница 356: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 36 ...
Страница 357: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37 ...
Страница 358: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37A ...
Страница 359: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 37B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 360: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 38 ...
Страница 361: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 39 ...
Страница 362: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 40 ...
Страница 363: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 41 ...
Страница 364: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 4 Direction 18000 7 2 42 ...
Страница 365: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 43 ...
Страница 366: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 44 ...
Страница 367: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 45 ...
Страница 368: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 46 ...
Страница 369: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 47 ...
Страница 370: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47A ...
Страница 371: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 47B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 372: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48 ...
Страница 373: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48A ...
Страница 374: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 2 Direction 18000 7 2 48B THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 375: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 49 ...
Страница 376: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 50 ...
Страница 377: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 51 ...
Страница 378: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 2 52 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 379: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 1 ...
Страница 380: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 3 2 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 381: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 1 ...
Страница 382: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 15 Direction 18000 7 4 2 ...
Страница 383: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 4 3 ...
Страница 384: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 7 4 4 ...
Страница 385: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 1 OVERTEMPERATURE SENSOR SCHEMATIC COMPUTER CABINET ...
Страница 386: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 6 2 OVERTEMPERATURE SENSOR UNIT COMPUTER CABINET ...
Страница 387: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 1 REMOTE CONTROL STATION POWER ...
Страница 388: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 7 7 2 REMOTE CONTROL STATION POWER ...
Страница 390: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 ii THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 402: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 1 12 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 411: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 5 ...
Страница 420: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 1 Direction 18000 8 3 16 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 428: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 4 8 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 462: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 24 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 472: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 3 Direction 18000 8 6 34 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 478: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 16 Direction 18000 8 7 6 THIS PAGE LEFT BLANK INTENTIONALLY ...
Страница 483: ...GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM Rev 19 Direction 18000 8 8 5 ILLUSTRATION A POSITIONING THE DETECTOR ...
Страница 487: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 3 ...
Страница 489: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 5 ...
Страница 490: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 6 ...
Страница 491: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 7 ...
Страница 492: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 8 ...
Страница 493: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 9 ...
Страница 494: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 10 ...
Страница 495: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 11 ...
Страница 496: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 12 ...
Страница 497: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 13 ...
Страница 498: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 14 ...
Страница 499: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 15 ...
Страница 500: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 16 ...
Страница 501: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 17 ...
Страница 502: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 18 ...
Страница 503: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 19 ...
Страница 504: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 20 ...
Страница 506: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 22 ...
Страница 507: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 23 ...
Страница 508: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 24 ...
Страница 509: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 25 ...
Страница 510: ...GE MEDICAL SYSTEMS CT 9800 WF RELEASE ERRATA Rev 17 Direction 18000 8 A 26 ...