Section 2-2 - General console requirements
Section 2-2
General console requirements
2-2-1
Environmental Requirements
2-2-1-1
Cooling
The cooling requirement for the LOGIQ S8 is 680 BTU/hr. This figure does not include cooling needed
for lights, people, or other equipment in the room.
Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
Table 2-1
System Environmental Requirements
Operational
Storage
Transport
Temperature
10 - 35
o
C /50 - 95
o
F
-10 - 50
o
C/14 - 122
o
F
-10 - 50
o
C/14 - 122
o
F
Humidity
30 - 80% (non-condensing)
10 - 90% (non-condensing)
10 - 90% (non-condensing)
Pressure
700 - 1060 hPa
700 - 1060 hPa
700 - 1060 hPa
Heat Dissipation
680BTU/hr
n/a
n/a
Содержание 5418099
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank ...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork ...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Страница 417: ......