Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7 - 81
2.)
FibroScan Report and Probe button do not work properly
If FibroScan measurement and another function are performed at the same time, FibroScan
functions occasionally do not work properly.
Example 1:
If the button on the FibroScan probe is pressed while FibroScan Report is being
generated, the button may not respond.
Example 2:
Do not press the button on FibroScan probe, when FibroScan Report is selected and
the system starts generating the report.
<WORKAROUND>
Do not press the button on FibroScan probe, when FibroScan Report is selected and the system
starts generating the report.
3.)
Error message: “FibroScan is not available. A FibroScan module is required”
The following error message may appear when starting FibroScan mode.
<RECOVERY>
Reboot the system.
7-7-17
Fibroscan Probe
Indicator of Fibroscan probe is blinking when select the probe in Fibroscan mode. It’s indicated the
probe is active.
Содержание 5418099
Страница 11: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL ix ZH CN KO ...
Страница 12: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL x ...
Страница 16: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL xiv This page was intentionally left blank ...
Страница 32: ...GE HEALTHCARE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 16 ...
Страница 176: ...GE HEALTHCARE PROPRIETARY TO GE DIRECTION 5394227 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL 3 98 Section 3 16 Paperwork ...
Страница 416: ...GE HEALTHCARE DIRECTION 5394227 REVISION 12 LOGIQ S8 LOGIQ E8 SERVICE MANUAL Index 4 ...
Страница 417: ......